RH406、RH406L、RH406H
Adaptive for
Philips MX4000, MX6000 and Access series CT
Neusoft Neu Viz Dual and Neu Viz series CT
Non OEM designedX-ray tube housing
Replacement for Dunlee X-ray tube model
CTR1720 CTR1725 CTR1740
Rcan’s intellectual property
High vacuum technology
Better bearing protection system
The technical parameters are adaptive with the original design.
Main technical parameters | |
Filtration of the tube housing | >1.5mmAl(70kV) |
High voltage socket | In accordance with GB/T 10151-2008 |
Heat capacity of the x-ray tube housing | 5.2MHU(3.7MJ) |
Cooling method | Forced cooling by fan oil |
Temperature switch | Switch off in 45~65℃. Switch on in 75~85℃. |
Leakage radiation under loading status | ≤0.88mGy/h |
Input power of the nominal anode | Input power of the small focus filamentnominal anode is 28kW. Input power of the big focus filamentnominal anode is 48kW. |
Maximum continuous heat dissipation of X ray tube housing | 6000W |
Weight | About 67 Kg |
Operation mode | For intermittent loading continuous operation,under conditionsof 3000W load, loading for 3 minutes, need 5 minutes intervals. CT manufacturers can set upproper operating parameters based on the heat capacity of the CT equipment. |
Power | Powered by a specific power supply, which is in accordance withGB 9706.1-2007 |
Driven parameters of the rotating anode | ||||
Model | RH406H | RH406 RH406L | ||
Item | Start | Operation | Start | Operation |
Power frequency(Hz) | 120 | 50 | ||
Voltage (V) | 220 | 80 | 220 | 85 |
Current (A) | 14 | 3 | 15 | 3.7 |
Acceleration time (s) | 10 | —— | 10 | —— |
Speed (RPM) | —— | 6800 | —— | 2700 |
Property of all brands mentioned above do not belong to Zhuhai RCAN Medical Equipment Co., Ltd
Zhuhai RCAN Medical Equipment Co., Ltd reserves the right to explain the above description.