RH406、RH406L、RH406H

rh406  

Adaptive for


Philips MX4000, MX6000 and Access series CT
Neusoft Neu Viz Dual and Neu Viz series CT
Non OEM designedX-ray tube housing

Replacement for Dunlee X-ray tube model


CTR1720  CTR1725  CTR1740

Rcan’s intellectual property

High vacuum technology

Better bearing protection system

The technical parameters are adaptive with the original design.

Main technical parameters
Filtration of the tube housing>1.5mmAl(70kV)
High voltage socketIn accordance with GB/T 10151-2008
Heat capacity of the x-ray tube housing5.2MHU(3.7MJ)
Cooling methodForced cooling by fan oil
Temperature switchSwitch off in 45~65℃.
Switch on in 75~85℃.
Leakage radiation under loading status≤0.88mGy/h
Input power of the nominal anodeInput power of the small focus filamentnominal anode is 28kW.
Input power of the big focus filamentnominal anode is 48kW.
Maximum continuous heat dissipation of X ray tube housing6000W
WeightAbout 67 Kg
Operation modeFor intermittent loading continuous operation,under conditionsof 3000W load, loading for 3 minutes, need 5 minutes intervals. CT manufacturers can set upproper operating parameters based on the heat capacity of the CT equipment.
PowerPowered by a specific power supply, which is in accordance withGB 9706.1-2007
Driven parameters of the rotating anode
ModelRH406HRH406   RH406L
ItemStartOperationStartOperation
Power frequency(Hz)12050
Voltage (V)2208022085
Current (A)143153.7
Acceleration time (s)10——10——
Speed (RPM)——6800——2700