RX526

Adaptive for


Adaptive for 16 slice CT of various brands
Philips MX16
United imagine uCT510
Neusoft Neu Viz16
ANKE ANATOM 16
MINFOUND ScintCare 16
SINO VISION Insitum 16
WDM TURBOTOM 1600
Non OEM designed X-raytube housing

Replacement for Dunlee X-ray tube


CTR2150

Rcan’s intellectual property
High vacuum technology
Better bearing protection system
The technical parameters are adaptive with the original design.
Main technical parameters
Filtration of the tube housing(IEC60522/1999) 4.0mmAL(70kV)
High voltage socket In accordance with GB/T 10151-2008
Heat capacity of the x-ray tube housing 5500KHU(4061kJ)
Cooling method Forced cooling by fan oil
Leakage radiation under loading status ≤0.88mGy/h
Input power of the nominal anode Input power of the small focus filamentnominal anode is 36kW
Input power of the big focus filamentnominal anode is 50kW
Maximum continuous heat dissipation of the X ray tube housing 6000W
Weight About 17.5 Kg
Operation mode For intermittent loading continuous operation,under conditionsof 6kW load, loading for 3 minutes, need 5 minutes intervals. CT manufacturers can set upproper operating parameters based on the heat capacity of the CT equipment.
Power Powered by a specific power supply, which is in accordance with GB 9706.1-2007
旋转阳极驱动参数
Item Start Operation
Power frequency(Hz) 105 105
Voltage (V) 480 90
Current (A) 9.0 1.0
Acceleration time (s) 20 ——
Speed (RPM) —— 6500
Property of all brands mentioned above do not belong to Zhuhai Rcan Vacuum Electron Co., Ltd.