RX406

Adaptive for


Philips MX4000, MX6000 and Access Series CT
Neusoft New Viz Dua and New Viz Twin Series CT
Non OEMdesignedX-ray tube insert

Replacement for Dunlee X-ray tube model


DU2000、DU2506、DU4006

Rcan’s intellectual property
High vacuum technology
Better bearing protection system
The technical parameters are adaptive with the original design.
Main technical parameters
Anode rotating speed ≥6800 r/min(120Hz)
Input power of the anode nominal 28KW
48KW
Maximum heat capacity of the anode 4000KHU(2840KJ)
Voltage of the nominal X ray tube voltage 140KV
Overvoltage 150KV
Nominal value of focus spot(YY/T0063) 0.4x0.7
0.6x1.3
Targetplate material Graphite target plate based on an alloy of tungsten, rhenium and molybdenum
Target plate angle
Target plate diameter 165mm
Exposure rate ≥0.03mC/kg.S
Inherent filtration(IEC60522/1999) 1.0mmAL(70kV)
Weight About 6.0 kg
Property of all brands mentioned above do not belong to Zhuhai Rcan Vacuum Electron Co., Ltd.
Zhuhai Rcan Vacuum Electron Co., Ltd. reserves the right to explain the above description.